Medical power supplies / KC-NT-005
Medical desktop power supply · 200 W series
External medical power supply with universal AC input and regulated DC output.
Medical2 × MOPPIEC / EN 60601-1IEC 60601-1-2IEC / EN 62368-1CE · catalogue marking

Photo of a series configuration. Mains inlet and DC connector depend on the selected configuration.
Technical data
- Reference
- KC-NT-005
- AC input
- 100–240 V AC nominal; 90–264 V AC range
- Line frequency
- 50–60 Hz nominal; 47–63 Hz range
- Input current
- 2.4–1.0 A
- Mains inlets
- C14, C18
- Dimensions L × W × H
- 173 × 67 × 32.1 mm
- Operating temperature
- −20 to +40 °C
- Cooling
- Natural convection
- Topology
- LLC
- Protections
- Short circuit, overvoltage, overcurrent, overtemperature
Available output configurations
| Keepcool reference | Output voltage | Output current | Output power | Enquiry |
|---|---|---|---|---|
| KC-NT-005-12V | 12 V DC | 16 A | 192 W | |
| KC-NT-005-15V | 15 V DC | 13.33 A | 200 W | |
| KC-NT-005-19V | 19 V DC | 10.5 A | 200 W | |
| KC-NT-005-24V | 24 V DC | 8.3 A | 200 W | |
| KC-NT-005-25V | 25 V DC | 8 A | 200 W |
Mains inlet and matching cord connector

IEC C14
Matching cord connector: IEC C13

IEC C18
Matching cord connector: IEC C17
Specify the mains inlet in your enquiry. We will coordinate the DC equipment connector, polarity and required cable length with you.
Medical standards and series ratings
- Medical safety
- IEC / EN / ANSI / AAMI ES60601-1
- EMC
- IEC 60601-1-2
- Additional safety standard
- IEC / EN / UL 62368-1
- Means of protection
- 2 × MOPP
- Primary–secondary dielectric withstand
- 4,000 V AC
- Load regulation
- ±5 % typical
- Ripple and noise
- ≤1 % Vp-p at full load
- Start-up time
- <3 s
- Hold-up time
- ≥10 ms at full load
- Rise time
- <50 ms
- Storage temperature
- −20 to +80 °C (packed)
- Operating altitude
- 5,000 m
Standards and CE marking refer to the series catalogue information. Suitability for your complete system is coordinated for the specific configuration and application.
Discuss your requirements →